Instruction Manual: EMP-3
NIC-TD-0000067-01
Nippon Instruments Corporation
2
This instrument does not have explosion-proof structure.
Before measuring liquefied natural gas or liquefied petroleum
gas, be sure to collect some into a Tedlar Bag. Please measure
them in non-explosion-proof areas.
Do not let liquid such as water splash the instrument. Never
operate the instrument with a wet hand. Otherwise, an electric
shock, a short circuit of an electronic circuit or a failure caused
by corrosion may result.
Do not disassemble or modify the product without permission.
Otherwise, a failure may result. For inspection or repair,
consult with our agent or service department.
Use the instrument after reading this document carefully.
Otherwise, a failure caused by an operational mistake may
occur.
Observe the range of the specification. If the instrument is used
out of the specification range, a failure may occur.
Do not give a shock to the instrument by dropping, throwing or
the like. The limit of fall height is 3cm.
Do not apply a strong external force to the screen by pushing
strongly, striking or the like.
Do not operate the instrument before the connection of the
probe is completed. Otherwise, fine particles will enter the
instrument and accurate measurements may become
impossible.
Select a probe suitable for a measurement position.
If any time-change component is not replaced, original
performance may not be demonstrated.
For data output, we recommend the supplied USB memory.
Although it is also possible to use a USB memory other than
the supplied one, data output may fail or data in the USB
memory may be damaged depending on a type.
Gas after a measurement is discharged from the outlet. When
measuring liquefied natural gas or liquefied petroleum gas, use
the instrument in a draft chamber or connect the outlet and
draft chamber or exhaust duct using a tube to discharge the gas.
Warning