Screen switching: Advance
Screen switching: Return
The menu screen reappears.
5.4.4.1 GLP
The GLP information can be checked and the counter can be reset after the replacement of
the ZERO filter.
When the recommended replacement time has been reached, an error message appears.
Request the replacement of the pump or lamp of us or our sales agent.
For the replacement of the ZERO filter, refer to “REPLACEMENT OF ZERO FILTER”.
Be sure to press “RESET” after the replacement. (Refer to “6.5.1 REPLACEMENT OF
When the free space of the micro SD card is insufficient, delete saved data. (Refer to
No.
Name
Description
1
Pump
Pump driving time (Hr). Recommended replacement time: 2000 hours.
2
Lamp.
Lamp operation time (Hr). Recommended replacement time: 2000 hours.
3
Total time
Total operation time of the instrument
4
Storage
Usage rate of the micro SD card (%)
5
Total Hg
Quantity of mercury adsorbed by the ZERO filter (ng)
6
RESET
Resetting of the GLP counter of the ZERO filter