Instruction Manual: EMP-3
NIC-TD-0000067-01
Nippon Instruments Corporation
34
(2)
Set the maximum and minimum of the concentration axis and the time duration of the
time axis. Press the numerical button of an item to be set.
(3)
Enter a value to be set and press “Enter”.
Presettable Range
Max
:
-9.9-2100 μg/Nm
3
Min: -10-2099.9 μg/Nm
3
*The presettable range change according to the each input value.
Time duration
:
1-100 min.
*Time duration must be greater than 10 times of the data saving interval.
Button
Description
Cancel
Cancellation of alteration
Clear
Deletion of all characters
Enter
Determination
BS
Deletion of one character
(4)
Press “OK” to return to the measurement screen.