6.4 Performance Testing
HSA42011
6-10
6.4.4 Measurement of gain accuracy
Connection
Connect an FRA and terminating resistor as shown in "
Figure 6-4 Measurement of Frequency
Characteristics
".
Setting
Set this product as follows:
Item Setting
Input A
Input impedance
50
Bias addition
OFF, dial 5.00
Polarity switching (INVT)
OFF
Terminating resistor
50
Set the FRA as follows:
Item Setting
Output Sine
wave
Frequency 400
Hz
Analysis ch1/ch2
Indication
logF
R
Test procedure
<1> Set the gain of this product and the output voltage of the FRA as shown in "
Table 6-1 Judgment
of HSA42011 Performance
".
<2> Turn on the outputs of this product and FRA.
<3> Make a measurement in the continuous mode.