2.4 Simplified Operation Check
This section describes the simplified operation check method, which may be used when this product is
newly purchased or after being stored for a long time. For performance checking, refer to
"6.4
Performance Testing
".
This product contains high-voltage parts. Never remove the cover.
All internal inspections of this product are to be performed only by service technicians qualified by
NF Corporation.
Required measuring instruments
To make an operation check, the measuring instruments listed below are required.
No load test is conducted since a load (terminating) resistor is not needed.
Table 2-1. Required Measuring Instruments
Measuring instrument
Characteristics
Signal generator
Frequency: 1
kHz
Waveform: Sine
wave
Output voltage:
0.5 Vrms or more (Load: 50
WF1947 1 CH 30 MHz multifunction generator
manufactured by NF Corporation, or equivalent
Oscilloscope
Frequency bandwidth:
20 MHz or more
Use an oscilloscope that can measure 100 V or higher by
using a 10:1 probe
Note
.
Voltmeter
AC and DC voltage measurement
Use a voltmeter that can measure 100 V or higher.
Note:
When connecting an oscilloscope for measurements, be sure to use a 10:1 probe.
!
WARNING
Содержание HSA42011
Страница 1: ...NF Corporation HIGH SPEED BIPOLAR AMPLIFIER HSA42011 INSTRUCTION MANUAL...
Страница 2: ......
Страница 3: ...DA00082229 001 HIGH SPEED BIPOLAR AMPLIFIER HSA42011 INSTRUCTION MANUAL...
Страница 4: ......
Страница 18: ......
Страница 30: ......
Страница 46: ......
Страница 80: ......
Страница 82: ......
Страница 84: ......
Страница 85: ......
Страница 86: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN...