3.4 Insulation breakdown voltages of input and output voltages
FRA5087
3-20
OSC
CH1
CH2
250Vrms
250Vrms
250Vrms
250Vrms
250Vrms
250Vrms
250Vrms
250Vrms
250Vrms
250Vrms
250Vrms
250Vrms
Fig. 3-3 Oscillation section-analysis section inputs isolation voltage specifications
(when accompanying BNC cable is used)
OSC
CH1
CH2
30Vrms
30Vrms
30Vrms
30Vrms
30Vrms
30Vrms
30Vrms
30Vrms
30Vrms
30Vrms
30Vrms
30Vrms
Fig. 3-4 Oscillation section-analysis section inputs isolation voltage specifications
(when a cable other than the accompanying cable is used)
WARNING
!
Do not apply excessive voltages between insulated signal terminals. You
could be suffered from electric shocks, if excessive voltages are applied
between these terminals leading to dielectric breakdown.
You could be suffered from electric shocks when you measure high voltage
circuit signals. Use accessory coaxial cables of the insulation type by all
means, so that you cannot directly touch metallic portions of the BNC
connectors at the analyzer input terminals.
Содержание FRA5087
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Страница 2: ......
Страница 3: ...DA00014460 004 FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Страница 4: ......
Страница 35: ...2 2 Mounting and installation FRA5087 2 7 Fig 2 3 Removal of foot stands...
Страница 40: ...2 6 Calibration FRA5087 2 12 Blank...
Страница 120: ...4 19 Condition display FRA5087 4 46 Blank...
Страница 134: ...5 6 Max min search function FRA5087 5 14 Blank...
Страница 158: ...7 2 Quick diagnosis FRA5087 7 8 Blank...
Страница 170: ...8 5 Performance test FRA5087 8 12 Blank...
Страница 190: ......
Страница 192: ......
Страница 193: ......
Страница 194: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima higashi Kohoku ku Yokohama 223 8508 JAPAN...