3.5 Examples of basic operations
FRA5087
3-26
3.5.3
Examples of basic measurement operations
This section describes setting procedures for basic measurement operations. The descriptions
assume that the setting status of the FRA5087 is the same as that at a delivery from factory or that
right after initialization.
a)
Example of SINGLE measurement
The SINGLE measurement is a measurement mode where only one time of measurement is
made at a fixed frequency. An example of measurement is described below with the
following measurement conditions:
・
Measurement frequency
: 1.0 MHz
・
Oscillator
: Amplitude=5 Vpeak and DC bias=1 V (the amplitude
to be set with the output being open)
The following parameters are unchanged from those at completion of initialization:
・
Oscillation waveform
: sinusoidal
・
Number of times of integration : 1 cycle
・
Delay time
: 0 cycle
・
Measurement mode
: CH1 and CH2
・
Analysis mode
: CH1/CH2
・
Display mode
: X-axis : logF
Y1-axis : dBR
Y2-axis :
1)
Connection with the SUT (system under test)
Since the measuring mode is CH1 and CH2, connect the FRA5087 and the system under
test (SUT) as shown in Fig. 3-12.
FRA5087
SUT
In
Out
BNC-T branch
connector
OSC
CH1 CH2
Fig. 3-12 Connection with SUT
Содержание FRA5087
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Страница 2: ......
Страница 3: ...DA00014460 004 FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Страница 4: ......
Страница 35: ...2 2 Mounting and installation FRA5087 2 7 Fig 2 3 Removal of foot stands...
Страница 40: ...2 6 Calibration FRA5087 2 12 Blank...
Страница 120: ...4 19 Condition display FRA5087 4 46 Blank...
Страница 134: ...5 6 Max min search function FRA5087 5 14 Blank...
Страница 158: ...7 2 Quick diagnosis FRA5087 7 8 Blank...
Страница 170: ...8 5 Performance test FRA5087 8 12 Blank...
Страница 190: ......
Страница 192: ......
Страница 193: ......
Страница 194: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima higashi Kohoku ku Yokohama 223 8508 JAPAN...