4.9 Amplitude compression
FRA5087
4-25
Use the menu [Measure][AMPLITUDE COMPRESSION] to set the amplitude compression.
The menu compression setting items are indicated in Table 4-4.
Table 4-4 Amplitude compression setting items
Items
Descriptions
[FUNCTION]
To be set at “ON” when you want amplitude compression.
[REF CHANNEL]
To be used to set the channel (CH1 or CH2) for level monitoring.
[REF LEVEL]
To be used set the reference level to be measured in the [REF
CHANNEL].
[OUTPUT LIMIT]
To be used to set the oscillator maximum output voltage.
[ERROR]
To be used to set the permissible difference between the [REF
LEVEL] set value and the measured data.
[RETRY TIMES]
To be used to set the maximum repeatable number of
measurement.
[CORRECTION FACTOR]
To be used set the voltage correction factor.
An example of amplitude compression is shown below where the voltage correction factor
[CORRECTION FACTOR] has been set at 70 %. It is also assumed that the current oscillator
setting voltage is 1 V and the reference level [REF LEVEL] of the oscillator is 2 V.
0
1
2
3
1
2
1V
V
0
1.7V
V
1
1.91V
V
2
1.973V
V
3
Number of measurements n
Oscillator
output voltage
Output reference voltage
V
G
=2V
V
n
=(V
G
-V
n-1
)×0.7+V
n-1
Fig. 4-17 Example of output correction (70 %)
Note that the oscillator output voltage is different from the amplitude set in the menu
[OSC][AMPLITUDE], even if you have disabled the amplitude compression function by setting
the [FUNCTION] at “off”.
!
CAUTION
Error message 15 may appear during amplitude compression. This is not a
malfunction.
See 7.1 error message.
Review [ERROR], [RETRY TIMES] and [CORRECTION FACTOR]
settings. Again conduct measurement.
Содержание FRA5087
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Страница 2: ......
Страница 3: ...DA00014460 004 FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Страница 4: ......
Страница 35: ...2 2 Mounting and installation FRA5087 2 7 Fig 2 3 Removal of foot stands...
Страница 40: ...2 6 Calibration FRA5087 2 12 Blank...
Страница 120: ...4 19 Condition display FRA5087 4 46 Blank...
Страница 134: ...5 6 Max min search function FRA5087 5 14 Blank...
Страница 158: ...7 2 Quick diagnosis FRA5087 7 8 Blank...
Страница 170: ...8 5 Performance test FRA5087 8 12 Blank...
Страница 190: ......
Страница 192: ......
Страница 193: ......
Страница 194: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima higashi Kohoku ku Yokohama 223 8508 JAPAN...