5.1 General
FRA5087
5-2
5.1 General
The PA-001-1231 is an optional added software package for the FRA5087 Frequency Response
Analyzer. The additional functions provided are impedance display, open and short correction,
maximum and minimum value search.
Since installation requires overwriting the FRA5087 internal software, this option must be
installed by the factory or a properly equipped dealer.
5.2 Features
5.2.1 Impedance display
Linear and logarithmic graphs display impedance, reactance, admittance, conductance and
susceptance. Current to voltage conversion coefficient (0 to 1.0E+6) can also be set.
5.2.2 Open and short correction
This function reduces systematic impedance measurement errors due cable and other factors.
Before measuring impedance, the terminals are first measured at open and short states and the
resulting data stored in memory. During actual measurement, the resulting open and short data
are shown graphically.
5.2.3 Max/min value search
The maximum or minimum value of the vertical axis parameters is automatically searched and the
computed value displayed.
Содержание FRA5087
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Страница 2: ......
Страница 3: ...DA00014460 004 FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Страница 4: ......
Страница 35: ...2 2 Mounting and installation FRA5087 2 7 Fig 2 3 Removal of foot stands...
Страница 40: ...2 6 Calibration FRA5087 2 12 Blank...
Страница 120: ...4 19 Condition display FRA5087 4 46 Blank...
Страница 134: ...5 6 Max min search function FRA5087 5 14 Blank...
Страница 158: ...7 2 Quick diagnosis FRA5087 7 8 Blank...
Страница 170: ...8 5 Performance test FRA5087 8 12 Blank...
Страница 190: ......
Страница 192: ......
Страница 193: ......
Страница 194: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima higashi Kohoku ku Yokohama 223 8508 JAPAN...