4.1 Overview of measurement and processing
FRA5087
4-3
4.1.1 Measurement mode
There are three (3) measurement modes available, i.e., (CH1, CH2), (CH1, OSC) and (OSC, CH2),
one of which is selected for the measurement. The measurement mode is set through the menu
[Measure][BASIC FUNCTION][MEASURE MODE].
Note: The measurement mode (CH1, CH2), for example, is sometimes expressed as (CH1 and
CH2) in this Instruction Manual.
CH1
CH2
O SC
D
F
T
V 1
=|V
1
|,θ
1
V 2
=|V
2
|,θ
2
CH1
CH2
O SC
D
F
T
V 1
=|V
1
|,θ
1
V 2
=|V
2
|,θ
2
a) CH1, CH2
b) CH1, OSC
CH1
CH2
O SC
D
F
T
V 1
=|V
1
|,θ
1
V 2
=|V
2
|,θ
2
c) OSC, CH2
Fig. 4-2 Measurement mode and internal connection
The measurement mode (CH1, CH2) is used for the three-port system and the measurement
modes (CH1, OSC) and (OSC, CH2) are used for the two-port system. Note, however, that the
two-port system measurement provides better results for the measurement mode (CH1, CH2),
since the effect of cables can be reduced in the two-port system measurement.
CH1
CH2
OSC
S.U.T
In
Out
FRA5087
Measurement mode: (CH1,CH2)
Fig. 4-3 Recommended measurement connection for two-port system
Содержание FRA5087
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Страница 2: ......
Страница 3: ...DA00014460 004 FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Страница 4: ......
Страница 35: ...2 2 Mounting and installation FRA5087 2 7 Fig 2 3 Removal of foot stands...
Страница 40: ...2 6 Calibration FRA5087 2 12 Blank...
Страница 120: ...4 19 Condition display FRA5087 4 46 Blank...
Страница 134: ...5 6 Max min search function FRA5087 5 14 Blank...
Страница 158: ...7 2 Quick diagnosis FRA5087 7 8 Blank...
Страница 170: ...8 5 Performance test FRA5087 8 12 Blank...
Страница 190: ......
Страница 192: ......
Страница 193: ......
Страница 194: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima higashi Kohoku ku Yokohama 223 8508 JAPAN...