8.5 Performance test
FRA5087
8-4
8.5 Performance test
This section describes, among principal test items, those items which can be tested without using
special/elaborate jigs or measurement equipment. Adjustment and/or repair is needed when you
have any test results that do not meet the specifications.
Ask NF Corporation for more detailed tests, calibrations and/or repairs.
8.5.1 Test equipment
The following test equipment and cables are required for performance tests:
Frequency counter: Accuracy: 1
10
-6
or better
Multi-meter (the following measurement shall be possible):
AC voltage:
Accuracy:
0.1 % for 100 mV - 10 V and 100 Hz - 10 kHz.
DC voltage:
Accuracy:
0.1 % for 100 mV - 10 V
Wideband multi-meter (the following measurement shall be possible):
AC voltage: Accuracy:
0.5 % for 100 mV - 10 V and 100 kHz - 200 kHz
Accuracy:
1 % for 100 mV - 10 V and 200 kHz - 1 MHz
Accuracy:
5 % for 100 mV - 10 V and 1 MHz – 15 MHz
Distortion meter:
0.1 % or better for full-scale
Others:
BNC-BNC coaxial cable, T-type divider, etc.
8.5.2 Pretest preparations
a)
Confirmation of test environment
Perform the test under the following conditions:
Ambient temperature:
+18 - +28 ºC Ambient humidity: 25 - 75 %RH
Power supply voltage:
90 - 132 VAC or 180 - 250 VAC
b)
Confirmation of functions/operations
Check and confirm important functions, before the test, in an abbreviated manner according
to
“2.5 Quick function checking”
.
c)
Warm-up
Leave the instrument for at least one (1) hour after turning the power on so that the internal
temperature becomes stabilized.
Perform/execute calibration by all means before the test. The performance of the FRA5087
is specified on the basis of the status right after calibration.
Содержание FRA5087
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Страница 2: ......
Страница 3: ...DA00014460 004 FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Страница 4: ......
Страница 35: ...2 2 Mounting and installation FRA5087 2 7 Fig 2 3 Removal of foot stands...
Страница 40: ...2 6 Calibration FRA5087 2 12 Blank...
Страница 120: ...4 19 Condition display FRA5087 4 46 Blank...
Страница 134: ...5 6 Max min search function FRA5087 5 14 Blank...
Страница 158: ...7 2 Quick diagnosis FRA5087 7 8 Blank...
Страница 170: ...8 5 Performance test FRA5087 8 12 Blank...
Страница 190: ......
Страница 192: ......
Страница 193: ......
Страница 194: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima higashi Kohoku ku Yokohama 223 8508 JAPAN...