4.5 Oscillator setting
FRA5087
4-10
4.5
Oscillator setting
4.5.1 Oscillator basic setting
Use the menu [OSC] to set the parameters, states, etc. pertaining to the oscillator.
[FREQUENCY]
This is used to set the oscillator frequency. This is primarily used for measurements at
fixed frequencies (SINGLE measurement or REPEAT measurement).
[AMPLITUDE], [DC BIAS]
These are used to set the AC output voltage amplitude (Amplitude) and the DC output
voltage (DC Bias) of the oscillator. Press the
AC/DC ON
key by all means, even if the
AC and DC lamps are lit on at the time of setting operation.
!
CAUTION
The output voltage will not change even if you change the output voltage
settings, unless you press the
AC/DC ON
key.
[STOP MODE]
One of the following is selected in the oscillator stop mode, when you press the
AC OFF
key:
[ZERO] : The oscillator stops with the AC output voltage becoming 0 V at the instant of
your pressing the key.
[HOLD] : The oscillator stops with the AC output phase being maintained at the instant
of your pressing the key.
[PHASE] : The oscillator stops at the time when the output phase becomes the same as
that set by [START/STOP PHASE] through time elapse.
Regardless of the setting mentioned above, the output voltage becomes 0 V when the
AC/DC
OFF
key is pressed.
The oscillator output voltages are shown in the figure as
“Fig. 4-7 Oscillator stop
mode”
, in which the
AC OFF
key is pressed first while the oscillator output is on, then
the
AC/DC ON
key is pressed.
Содержание FRA5087
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Страница 2: ......
Страница 3: ...DA00014460 004 FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Страница 4: ......
Страница 35: ...2 2 Mounting and installation FRA5087 2 7 Fig 2 3 Removal of foot stands...
Страница 40: ...2 6 Calibration FRA5087 2 12 Blank...
Страница 120: ...4 19 Condition display FRA5087 4 46 Blank...
Страница 134: ...5 6 Max min search function FRA5087 5 14 Blank...
Страница 158: ...7 2 Quick diagnosis FRA5087 7 8 Blank...
Страница 170: ...8 5 Performance test FRA5087 8 12 Blank...
Страница 190: ......
Страница 192: ......
Страница 193: ......
Страница 194: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima higashi Kohoku ku Yokohama 223 8508 JAPAN...