4.1 Overview of measurement and processing
FRA5087
4-2
4.1 Overview of measurement and processing
This section describes the flow of data processing from measurement up to data display in the
FRA5087. (See
“Fig. 4-1 Flow diagram of measurement and processing”
.)
CH1
CH2
O SC
Measurement
Measurement mode
Weighting factor
Calibration data
Equalizing
memory
Analysis
Analysis mode
Display
Display mode
(measurement
processing)
(Input
weighting)
(Self error
compensation)
(Error compensation
for measurement
system)
(Analysis
processing)
(Display in
graphs, etc.)
V1
V2
Buffer memory
Fig. 4-1 Flow diagram of measurement and processing
Measurement processing: This phase determines the connection of input terminals and of the
internal circuit. The measurement is done through DFT (discrete
Fourier transform).
Self error compensation: Self error compensation is made through the use of calibration data
obtained by calibration. Calibration can be done either at the time of
switching power supply on, or through the menu [Calib.].
Error compensation for measurement system:
The characteristics of various measurement equipment connected
externally to the FRA5087 are measured as measurement system
errors. The errors can be canceled at actual measurement (through
the equalizing capability).
Input weighting: The measurement data of CH1 and CH2, individually, can be weighed by
arbitrary individual factors of the range between 0 and 1.0E+6 through the
menu [Input][WEIGHTING FACTOR].
Buffer memory: The measured and error compensated data are stored in the temporary buffer
memory.
Analysis processing: Analysis is made in the measurement process of measurement through
error compensation phases, and the original measurement data are
converted to the data to be displayed.
Display processing: The analyzed data are displayed in the form of graphs on the LCD screen
according to the display format that has been set.
Data obtained in the sweep-mode measurement are once stored in the buffer memory while data
analysis and display are also being done. The data read out from the buffer memory can be used
for analysis and display format conversion without remeasurement.
Содержание FRA5087
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Страница 2: ......
Страница 3: ...DA00014460 004 FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Страница 4: ......
Страница 35: ...2 2 Mounting and installation FRA5087 2 7 Fig 2 3 Removal of foot stands...
Страница 40: ...2 6 Calibration FRA5087 2 12 Blank...
Страница 120: ...4 19 Condition display FRA5087 4 46 Blank...
Страница 134: ...5 6 Max min search function FRA5087 5 14 Blank...
Страница 158: ...7 2 Quick diagnosis FRA5087 7 8 Blank...
Страница 170: ...8 5 Performance test FRA5087 8 12 Blank...
Страница 190: ......
Страница 192: ......
Страница 193: ......
Страница 194: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima higashi Kohoku ku Yokohama 223 8508 JAPAN...