4.12 Auto-sequence
FRA5087
4-32
4.12.2 Executing key sequence
Select [RUN MODE] in the menu [AutoSeq.][MODE]. Then, press the
AUTO SEQ
key to start
execution of recorded key sequence operations with the PLAY lamp turned on, which indicates
that the recorded key sequence is being run. Press the
GPIB LOCAL
key to terminate the
execution midway during running the key sequence operation.
During key sequence, the operation proceeds as follows.
・
OSC (oscillator) ON/OFF (including SLOW ON/OFF)
・
Frequency sweep measurement, SINGLE measurement
・
Calibration
・
USB
flash drive
access
・
Hard-copying output
・
Calculations
!
CAUTION
If an error occurs midway during execution of the key sequence (amplitude
compression error “15 Amplitude compression failure” excepted, however),
the execution terminates at that point.
If an overload occurs midway during execution of the key sequence, the
execution is terminated only when either one of [SWEEP STOP] or
[OSCOFF] in the menu [Input][ACTION] is [ON].
If a key sequence involving a REPEAT measurement is executed, the
REPEAT measurement is repeated permanently without going to the next
execution. Therefore, do not record/store/execute any key sequence
involving REPEAT measurement. Press the
GPIB LOCAL
key to terminate
the REPEAT measurement in the key sequence.
When any access, control or inquiry is made from the GPIB (i.e., when the
status turns to REMOTE) during execution of key sequence, the execution is
terminated.
During execution of key sequence operations, all key entries are ignored
except for the
GPIB LOCAL
key.
If the work area memory cannot be secured at the time of starting execution,
the execution does not start with an error message (“Memory overflow”)
displayed.
Содержание FRA5087
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Страница 2: ......
Страница 3: ...DA00014460 004 FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Страница 4: ......
Страница 35: ...2 2 Mounting and installation FRA5087 2 7 Fig 2 3 Removal of foot stands...
Страница 40: ...2 6 Calibration FRA5087 2 12 Blank...
Страница 120: ...4 19 Condition display FRA5087 4 46 Blank...
Страница 134: ...5 6 Max min search function FRA5087 5 14 Blank...
Страница 158: ...7 2 Quick diagnosis FRA5087 7 8 Blank...
Страница 170: ...8 5 Performance test FRA5087 8 12 Blank...
Страница 190: ......
Страница 192: ......
Страница 193: ......
Страница 194: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima higashi Kohoku ku Yokohama 223 8508 JAPAN...