8.5 Performance test
FRA5087
8-11
8.5.9 Analyzer measuring error frequency response
Frequency dependence of the CH1/CH2 measurement value at 10 mVpeak output is tested.
a)
Setting
FRA5087 oscillator
FRA5087 analyzer
Sweep upper limit frequency
10 MHz
Integration
50 cycles
Sweep lower limit frequency
10 Hz
Measuring mode
CH1, CH2
Output waveform
Sinewave
Analyzing mode
CH1/CH2
Output voltage
Display mode
X axis : logF
AC
10 mVpeak
Y1
axis
: dB
R
DC bias
0 V
X2 axis :
b)
Connection
FRA5087
BNC-BNC
coaxial cables
BNC
T-type divider
OSC
CH1 CH2
c)
Procedure
Use FRA5087 for sweep measurement.
Read the absolute maximum measured results for dBR and
θ
for frequency ranges up to 20
kHz, 500 kHz, 2.2 MHz and 10 MHz from the LCD screen.
d)
Judgment
Frequency range
Measured value
Tolerance
10 Hz to 20 kHz
dB
-0.05to +0.05 dB
deg
-0.3 to +0.3 deg
20 kHz to 500 kHz
dB
-1.0 to +1.0 dB
deg
-0.5 to +0.5 deg
500 kHz to 2.2 MHz
dB
-1.0 to +1.0 dB
deg
-2.0 to +2.0 deg
2.2 MHz to 10 MHz
dB
-2.0 to +2.0 dB
deg
-5.0 to +5.0 deg
Содержание FRA5087
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Страница 2: ......
Страница 3: ...DA00014460 004 FREQUENCY RESPONSE ANALYZER FRA5087 INSTRUCTION MANUAL...
Страница 4: ......
Страница 35: ...2 2 Mounting and installation FRA5087 2 7 Fig 2 3 Removal of foot stands...
Страница 40: ...2 6 Calibration FRA5087 2 12 Blank...
Страница 120: ...4 19 Condition display FRA5087 4 46 Blank...
Страница 134: ...5 6 Max min search function FRA5087 5 14 Blank...
Страница 158: ...7 2 Quick diagnosis FRA5087 7 8 Blank...
Страница 170: ...8 5 Performance test FRA5087 8 12 Blank...
Страница 190: ......
Страница 192: ......
Страница 193: ......
Страница 194: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima higashi Kohoku ku Yokohama 223 8508 JAPAN...