![NeoMedLight BiliCocoon Bag System Скачать руководство пользователя страница 33](http://html1.mh-extra.com/html/neomedlight/bilicocoon-bag-system/bilicocoon-bag-system_instructions-for-use-manual_719135033.webp)
Instructions for Use :
BiliCocoon™ Phototherapy System
IFU-BCC-SY-EnFr-2017-10-v13
30
EN
1.
Remove the 4
screws (part
numbers: H10,
CBLX, M3x10, ISO
14583, Inox 2) with a
Torx adapted
screwdriver and
then the back cover
of the Light Box.
The figure on the
right shows the
location of the
screws:
2.
Locate
the
Fuse,
shown
in
the
pictures on the right:
3.
Remove the fuse while taking care not to damage the electronic components nearby.
4.
Replace the fuse with a new one. Use only the part number “SCHURTER 0001.1013”, which is
provided in the service kit. Do not use any other type of fuse.
5.
Verify the irradiance of the system before using it