![NeoMedLight BiliCocoon Bag System Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/neomedlight/bilicocoon-bag-system/bilicocoon-bag-system_instructions-for-use-manual_719135023.webp)
Instructions for Use :
BiliCocoon™ Phototherapy System
IFU-BCC-SY-EnFr-2017-10-v13
20
EN
Warning
5. Perform a visual inspection. If needed repeat the previous instruction until there is no
more soil observed.
Disinfection
6. Use one moistened clean cloth with a suitable disinfectant agent, or pre-saturated
wipe adapted to medical device cleaning (refer to the list of allowed disinfectant
agent). Wipe the device from the plug, to the pad to thoroughly wet the surface. Give
special attention to hard to reach areas of the device. Repeated use of the product
may be required to ensure that the surface remains visibly wet for a minimum of 3
minutes at r
oom temperature (+15°C/+25˚C). Take care when cleaning the Pad
connection to avoid damaging the optical fibers.
7. Moisten another clean soft cloth with water and wipe the Pad taking care to remove
the disinfectant residues.
8. Allow the device to air dry thoroughly
9. Before use, make sure the Pad is completely dry.
5.D No Cleaning for the Disposable
Caution
The
BiliCocoon™ Disposable is a SINGLE USE component, it is not designed to be reused,
disinfected or cleaned.
5.E Failures and Possible Resolutions
Event
Causes
Actions
Operation indicator
(blue
light strip)
– off
No therapy session is
configured
Operation indicator
(blue
light strip)
– flashing
The Pad is disconnected or
not properly connected. Even
if a session has been
configured, it’s not active.
Check that the Pad is properly connected to the
Light Box.
Reconnect the Pad if necessary.
The operation indicator will turn a continuous
blue to indicate that the session is active.
If the blue indicator light continues to flash,
contact the technical support.
LED failure indicator
–
continuous light
Session configured but one or
more LEDs has failed
The session has stopped and
cannot be restarted.
Contact technical support.
Overheating indicator
–
continuous light
Fan failure
The session has stopped
Contact technical support.
Switch the Light Box OFF and ON
If the fan failure indicator light is on, put the
Device out of service.
If the fan failure indicator does not light up and
the session does not restart after 5 seconds:
•
Turn OFF the Light Box and wait 15
min before switching it ON again.