![NeoMedLight BiliCocoon Bag System Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/neomedlight/bilicocoon-bag-system/bilicocoon-bag-system_instructions-for-use-manual_719135020.webp)
Instructions for Use :
BiliCocoon™ Phototherapy System
IFU-BCC-SY-EnFr-2017-10-v13
17
EN
in the picture; the other
surface is folded on the
infant’s chest.
XXX
4.E Performing a Treatment
Precautions
Warning
Before beginning the treatment check the installation and the Light Box / Pad connection.
When the Device is switched ON but the Pad is not connected, do not look at the Pad
Connection port.
Warning
During the session or at its end, the temperature of the Connection surface which is in contact
with the Light Box can be high. There is a risk of burning
– see section 5. Take care in
handling the Device.
Warning
Keep the Light Box steady with one hand when disconnecting the Pad through its
Connection. Pulling the Connection without keeping the Light Box may lead the Light Box to
fall, be damaged and harm users and other people.
Setting a Session
1. Turn ON the Light Box using the ON/OFF black
switch (1)
2. Set the session time (HH:MM) using the
+
and
–
controls (2). If a session was previously set refer
to the following section.
3. The session automatically begins in 5 seconds
with the Device emitting blue light through the disposable. Check that the Pad and the Light Box are well
connected: the Blue line feature light (3) should be continuous. In case the light flashes or is off, refer to the
table "Failures and possible resolutions".
4. When the session starts all the controls become inactive. To reactivate them, switch OFF and ON the
Device.
5. The session will stop automatically when the session time will be 00:00.
Resetting a Session or Changing the Settings of a Session
1. Switch OFF (1)
2. Switch ON (1)
3. To set the session time to zero press the RESET (4) button within 5 seconds from the switch ON
4. Set the session time (HH:MM) using the
+
and
–
controls (2) within 5 seconds after powering up.
Interrupting a Session
1. Switch OFF (1) to interrupt the session.
2. Switch ON (1) to restart a session: The system displays the last session time.