![NeoMedLight BiliCocoon Bag System Скачать руководство пользователя страница 32](http://html1.mh-extra.com/html/neomedlight/bilicocoon-bag-system/bilicocoon-bag-system_instructions-for-use-manual_719135032.webp)
Instructions for Use :
BiliCocoon™ Phototherapy System
IFU-BCC-SY-EnFr-2017-10-v13
29
EN
1.
Remove the 4
screws (part
numbers: H10,
CBLX, M3x10, ISO
14583, Inox 2) with a
Torx adapted
screwdriver and
then the back cover
of the Light Box.
The figure on the
right shows the
location of the
screws:
2.
Locate the dust filter
and its cover, shown
in the pictures on the
right:
3.
Remove the cover, to do so a small flat screwdriver might be used to push through the holes
situated next to the screw holes. Note that the screws that hold the Fan do not have to be
removed in order to replace the filter.
4.
Replace the filter behind the cover with a new one. Use only the part number “SANYO DENKI
109-
1001F20”, which is provided in the service kit. Do not clean and/or reuse the old filter.
5.
Place the cover over the new filter by applying a small pressure.
6.
Verify the irradiance of the system before using it
A.5 Fuse Replacement
Warning
Before starting the operation assure that the external
power supply is not plugged to the
Light Box.
This operation should
only be performed by the authorized technical
staff
of the hospital
or the supplier.
Replacement Procedure: