Instructions for Use :
BiliCocoon™ Phototherapy System
IFU-BCC-SY-EnFr-2017-10-v13
26
EN
9 WARRANTY
The Light Box and the Pad are covered by a 1-year warranty.
ANNEX A: Service Manual
A.1 Purpose
This Service Manual describes the
maintenance operations to be performed on the BiliCocoon™ Phototherapy
System exclusively by the authorized technical staff of the hospital or the supplier.
Applicable part number:
BiliCocoon™ Light Box – BCC010
List of Interchangeable or Detachable Parts (excluding Pad and disposable):
Part
Part Number
Power
Supply
MEGMEET MANGO100-12B
Power Cord
EU : Qualtek 360007-01
UK : Cabcon UKA03.H02.C07.0200.B
US : Schurter 6010.5274
AU/NZ : IEC-C7 2m AC-Cord Blk 2-Pin Plug to Iec320-C7 - K3762
Service Kit
BCC010-SK10
•
Fuse
: SCHURTER 0001.1013 (8A/250V/FAST/Breaking Capacity : H (1500A) and
size: 5x20mm)
•
Dust Filter
: SANYO DENKI 109-1001F20
A.2 Maintenance Operations
The following sections detail the preventive and corrective maintenance operations to be carried out. No other
operation is authorized on the Device.
A.2.1 Failures and Corrective Operations
Warning
The failure to perform the required corrective maintenance by authorized technical staff of
the hospital or of the supplier, may cause failures or malfunctions which may in turn cause
risks to the patient, user, third parties or properties.
Warning
DO NOT ATTEMPT TO REPLACE
the LEDs at their end of life, neither by identical LEDs
nor by a different component. CONTACT YOUR SUPPLIER.
The following table lists the failures that may occur and the actions to be undertaken. If the problem persists
after performing the maintenance operation, contact the supplier.
Observation
Failure
Actions
LED failure indicator
–
continuous light
LED failure
Contact the supplier