Instructions for Use :
BiliCocoon™ Phototherapy System
IFU-BCC-SY-EnFr-2017-10-v13
9
EN
Important
The Pad emitting surface should be stored flat, do not ever bend or stretch the emitting
surface optical fibers. Do not ever bend the optical fibers at right angles.
The cable must not be coiled more than one and a half turns. A manipulation that does not
follow the recommended instructions may damage the optical fibers and affect the Pad
lighting performances or lead to a risk to the patient's health.
Transport and Storage Conditions for Packaged Product
Warning
INDOOR storage only.
Temperature range: from -25 to 50°C
j
Relative humidity range: from 10 to 90%, without condensation
Atmospheric Pressure range: from 504 to 1060 hPa
k
3.B Light Box Features
The BiliCocoon™ Light Box is an electronic device incorporating LEDs and emits between 430 and 490 nm.
This emission range corresponds to the absorption spectrum of bilirubin, with this light leading to bilirubine
conversion and elimination without the liver intervention.
The Light Box includes an ergonomic handle which allows an intuitive grasp.
1
Ergonomic
handle
2
Front panel
with display and user controls
3
Connection
Port with the lenses which transfer the blue light to the Pad optical fibers
4
Side
ventilation (air inlets)
:
DO NOT OBSTRUCT
5
Back
ventilation (air inlets)
with dust filter (inside):
DO NOT OBSTRUCT
6
Power
socket (to be connected to the external power module)
7
USB Port
NOT OPERATIONAL
, DO NOT USE. The symbol “CAUTION” is
close to the USB port to recall not to use this port.
Controls
Controls can be used while the user is wearing medical gloves.
j
-13
– 122 °F
k
Equivalent Altitude : 0
– 5500 m (0 - 18044,62 feet)