Instructions for Use :
BiliCocoon™ Phototherapy System
IFU-BCC-SY-EnFr-2017-10-v13
21
EN
•
If the session does not restart after 5
seconds, put the Device out of service.
Overheating indicator
–
flashing light
Overheating
The session is stopped.
Verify that the ambient temperature does not
exceed the maximum operating temperature
(35°C
r
). In case of high ambient temperature,
move the system to an environment with a
temperature
below
35°C
and
let
the
temperature stabilize for 20 min.
Verify that the air inlets are not blocked.
If the air inlets are not blocked:
-
Contact the technical service
-
Switch OFF and ON the Light Box
-
If the session does not restart after 5
seconds, turn OFF the Light Box and wait
15 min before switching it ON again.
-
If the session does not restart after 5
seconds, depending on the urgency of the
treatment replace the phototherapy system
with one that works.
If the air inlets are blocked:
-
Remove anything that is obstructing the air
inlets
-
Switch the Light Box OFF and ON
-
If the session does not restart after 5
seconds, switch OFF the Light Box and
wait for 15 min before switching it ON
again.
-
If the session does not restart after 5
seconds, depending on the urgency of the
treatment replace the phototherapy system
with one that works.
-
In that case contact the technical support
The RESET or the session
time setting buttons do not
work
Defective Light Box
Do not use the Device.
Put the Device out of service.
Contact the technical support.
The Pad is cracked or
damaged
Defective Pad
Do not use the Pad or the Light Box the Pad is
paired with
Contact the technical support.
The Connection is cracked
or damaged
Defective Connection
Do not use the Pad or the Light Box the Pad is
paired with
Contact the technical support.
The Connection does not fit
or does not properly
connect with the Light Box
Defective Light Box or
defective Connection
Do not use the Device.
Put the Device out of service.
Contact the technical support.
The Light Box shell is
cracked or damaged
Defective Light Box
r
≈ 95°F