The blue light emission is stable from the session start: irradiance measurements can be taken as soon as
the phototherapy session is active.
1. Remove the Receptor Cap (1)
2. Switch On the Radiometer using the ON/OFF switch (2) on the side of the device as in the picture. Check
that the Hold-Run (3) is released (i.e. the display reading is not frozen). In case the display is frozen push
the Hold-Run to release it.
3. Measure the blue light from the environment while the
Device is turned off. Place the radiometer with the Light
receptor dome (4) oriented upward as in the picture on the
right. The displayed value should be lower than
1 μW∙cm
-
2
∙nm
-1
.
4. Connect the Pad to the Light Box, switch the Device on and launch a test session. The Pad now emits
blue light. See section 4.E.
5. To make a measurement place the light receptor dome (4)
in contact with the surface as in the picture
– in this
position the user looks at the back of the radiometer.
6. Place the light receptor dome on the measurement point
n° 1 as in the picture
7. Leave the light receptor dome in this position for few
seconds.
8. Press the Hold-Run (3) which freezes the measurement displayed on the screen.
9. Turn the radiometer, read and note the value displayed on the screen (5).
10. Push the Hold-Run (3) again to unfreeze the displayed measure.
11. Repeat steps 5 to 10 for the 7 other measuring points as shown in the pictures.