6 Maintenance and Troubleshooting
6.5 Troubleshooting
166
HORIBA
6.5.2 More troubleshooting
This section explains how to respond to various
symptoms of trouble that are not indicated by an error
number.
●
Nothing shows up on the display when the power is
turned ON
●
The indicated value fluctuates
When there is a problem with the electrode...
Cause
How to solve problem
No batteries
Place batteries in the meter.
The batteries
are loaded with
the poles
reversed.
Re-insert the batteries with the
poles correctly oriented.
The battery
voltage is low.
Remove the old batteries and
correctly insert new dry-cell
batteries.
Or connect the unit to the optional
AC adapter.
Cause
How to solve problem
The responsive
membrane is
dry or dirty.
Wash the responsive membrane.
The responsive
membrane is
damaged or
worn out.
Replace the electrode.
There are air
bubbles on the
electrode.
Shake the electrode to remove the
air bubbles.