2 Taking Measurements
2.3 Measurement modes
D-52/53/54/55
23
●
Auto Hold Measurement mode
Auto Hold Measurement mode maintains the display of
the value measured when the meter automatically
judges that the measured value has stabilized. Press the
MEAS key with the initial screeen displayed to make
HOLD blink on the display. When the measured value
becomes stable,
HOLD will stop blinking and remain
displayed, and the measured value will remain
displayed. To clear the hold status or
stabilized value
(when
HOLD is blinking), press the MEAS key.
Criteria for judging stability
Blinks
Displayed
pH, ORP, ION
measurement
:
Within ±1 mV variance in potential
after 10 seconds
Conductivity
measurement
:
Within ±3-digit variance after 10
seconds
DO measurement
:
Within ±3-digit variance after 10
seconds
Temperature
measurement
:
Within ±2ºC variance after 10
seconds