2 Taking Measurements
2.9 Dissolved oxygen (DO) measurement (D-55)
D-52/53/54/55
57
●
Air calibration
To achieve correct measurements, the pH meter must
be calibrated prior to taking measurements with it.
The D-50 Series pH meter can be calibrated using a
simple one-point air calibration and, when highly precise
measurement is required, using a two-point standard
solution calibration. This section explains the general air
calibration.
If a higher level of precision is required, refer to “
Calibrating with standard solution” page 61.
For greater measurement precision, it is necessary to
correct for air pressure.
Air-pressure correction
Air pressure is set to 1013 hPa, as the default. To
change this setting, refer to “3.4.13 DO atmospheric
pressure compensation setting (D-55 )” page 89.
1.
Remove any liquids from the membrane at the
tip of the electrode by either drying it or wiping
away the liquid with soft tissue paper, making
sure not to scratch the membrane.
2.
Press the CAL key while in the DO
Instantaneous Value Measurement mode, to
select the Calibration mode.