2 Taking Measurements
2.9 Dissolved oxygen (DO) measurement (D-55)
62
HORIBA
3.
Wash the electrode with tap water, and then
immerse it in the standard solution.
Mix the sample at a constant speed (1000 – 1500 rpm)
during measurement, using magnetic stirrer. When the
sample temperature rises due to the stirrer, use a
temperature bath.
With field-use electrodes, measure at a constant flow
speed (about 30 cm in 2 seconds).
4.
Press the CAL key.
“HOLD” will blink until the reading stabilizes.
While “HOLD” is blinking
To cancel calibration: Clear the hold by pressing the CAL
key, again.
To fix the calibration value: Fix the value using the
ENTER key.
After the readout value stabilizes, the HOLD is
displayed to indicate that the calibration is
completed.
Standard
solution
When the span
standard solution
is used:
When the zero
standard solution
is used: