2 Taking Measurements
2.5 Measuring pH
D-52/53/54/55
29
2.
Press the CAL key to start calibration.
The measured value will be displayed, and
HOLD will blink until the reading stabilizes.
When the value stabilizes,
HOLD will stop
blinking and the calibrated value will be displayed.
The
bottle mark will be displayed, indicating that
calibration was conducted with pH7 standard
solution
To stop the calibration:
Press the CAL key while the HOLD mark is blinking.
To establish the calibration:
Press the ENTER key while the HOLD mark is blinking.
To redo the calibration:
Press the CAL key after the HOLD mark is displayed.