4 RS-232C communications
4.4 Data request commands and responses
116
HORIBA
●
Ion calibration history request command and response
(D-53 )
Request command format
pH meter response format
・
The format is of fixed length. When there is no data,
enter a [SP].
・
The latest calibration date and time are output.
Channel number, 1-digit, 1: 1CH 2: 2CH
Command code
Header
Asymmetric potential (5-digit value including decimal
point and sign)
Temperature setting 1-digit, 0: ATC, 1: MTC
Calibration result 0: good; 1: check; 2: error
Calibration points 1-digit, 0-3
Channel number 1-digit, 1: CH1, 2: CH2
Header
First calibration data
Second calibration data
Third calibration data
Slope (sensitivity), potential (5 digits including decimal point
and sign)
*Potential is output for 1-point and 2-point calibration data.
Year, 4-digit (A.D.)
Month, 2-digit, 01 - 12
Month, 2-digit, 01 - 12
Minutes, 2-digit, 00 - 59
Hours, 2-digit, 00 -23
Seconds, 2-digit, 00-59
Calibration solution value
(5-digit value including decimal point and sign)
Temperature
(5-digit value including decimal point and sign)
Sub unit 0: none; 1:
μ
; 2: m
Unit 0: mol/L, 1: g/L
Potential
(5-digit value including sign)