6 Maintenance and Troubleshooting
6.5 Troubleshooting
164
HORIBA
●
ERR No.09 Printer error
Explanation
If a problem occurs with the printer, turn OFF the power
to the meter, perform the appropriate measure below,
and turn the power to the meter back ON.
●
ERR No.10 Data memory over
Explanation
The number of data items has exceeded the limit of the
memory.
Cause
How to solve problem
The printer
paper is
jammed.
Remove the jammed paper.
There is no
printer paper.
Load the printer with paper.
There is a
problem with the
printer
connection.
Reconnect the printer after making
sure there is nothing wrong with the
connector parts.
The printer is
defective.
Replace the printer.
Cause
How to solve problem
Memory over
Delete data stored in the memory
after confirming their contents.