6 Maintenance and Troubleshooting
6.1 pH (ORP) electrode maintenance
D-52/53/54/55
139
6
Maintenance and Troubleshooting
This chapter explains how to perform daily meter
maintenance and how to deal with error messages.
Daily maintenance is vital in assuring accurate measurement
and preventing breakdowns before they occur. Maintenance of
the electrodes is especially important; if ignored, various
problems and erroneous measurements may result. This
meter is equipped with a convenient error message function. If
an error message is displayed, be sure to take
appropriate action.
6.1
pH (ORP) electrode maintenance
Maintain your electrodes by referring to the following
information or to the operation manuals for the
electrodes.
The following explanation is for pH electrodes (9621-
10D) ORP electrodes should be cared for in the same
manner.
●
Maintenance after daily use
After taking measurements, wash the electrode using
pure water (de-ionized water), wipe off the water from
the electrode with filter paper or tissue paper, and store it
with its cap on.
Injury warning
Glass fragments can cause injury.
The outer tube of the electrode and the
tip of the electrode are made of glass.
Use care not to break them.
Caution