6 Maintenance and Troubleshooting
6.5 Troubleshooting
172
HORIBA
●
Measurements are not repeatable
●
When the printer will not print even though it is
connected
Check the following points:
・
Is the printer turned ON?
・
Has a printer error occurred?
・
Is there printing paper? Has the paper jammed?
・
When running a test print according to the manual,
does it print out correctly?
Cause
How to solve problem
Some effects of
the sample
solution
The pH or other properties of the
sample solution may have changed
over time, making repeatability
poor.
The responsive
membrane is
dry or dirty.
Wash the responsive membrane.
There is not
enough
reference
solution or it is
dirty.
Replace the reference solution with
new solution.
The responsive
membrane is
cracked or worn
out.
Replace the electrode.