2 Taking Measurements
2.5 Measuring pH
D-52/53/54/55
31
The example of calibration at second point has
explained the calibration process using the order from
pH 7 to pH 4. However, the calibration order of the
standard solutions can be arbitrarily chosen.
Electrode status
You can check the status of the electrode after
calibration.
Item
Description
,ERR
Not displayed
The electrode is in good condition.
Electrode sensitivity is from 93% to 100%.
Blinking
Electrode sensitivity has dropped to the
level of 90% to 93%.
・
Make sure that you are using the right
standard solution.
・
Clean the electrode.
Washing the electrodes” page 142
Displayed
Electrode sensitivity has dropped to the
level of 85% to 90%.
“
ERR No.05 Electrode sensitivity error
ERR No.04
Asymmetrical potential error
“
ERR No.04 Asymmetric potential
ERR No.05
Sensitivity error
“