2 Taking Measurements
2.7 Ion measurement (D-53)
D-52/53/54/55
41
●
Standard solution calibration
Calibrate the pH meter using a standard solution with a
known concentration.
Selecting the ions to be measured
The ION to be measured is set using the load count
(charge quantity).
The ION slope is set at +1 as the default setting.
Refer to “3.4.9 Ion slope setting (D-53)” page 86
Units
Units are set at g/L as the default setting. To change the
units to mol/L, refer to “3.4.8 Ion unit setting (D-53 )”
page 85.
Calibration at first point
1.
Press the CAL key while in the ION
Instantaneous Value Measurement mode, to
select the Calibration mode.
2.
Wash the electrode well with pure (de-ionized)
water, and then wipe with filter paper or tissue
paper.
Do not touch or scratch the responsive membrane on
the ION electrode.
3.
Place the tip of the electrode in the standard
solution beaker.