7 Reference
7.4 Conductivity measurement
190
HORIBA
7.4
Conductivity measurement
●
Electrode sensitivity check
The cell constant of a conductivity electrode may vary,
depending on the sample solution. Check the cell
constant by measuring conductivity, using the following
solutions, at least once every three months.
Prepare the potassium chloride standard solution (KCl
0.01 mol/L) using the procedure below.
In addition, if an error of 5% of more compared to the
above values occurs, calibrate the cell constant (See
page 52).
●
Preparing potassium chloride standard solution
How to prepare solution
Dry the potassium chloride powder (“superior quality”
commercial potassium chloride or better) for two hours,
at 105 ºC, then cool it in a desiccator. Measure out the
above-listed amount of potassium chloride into a beaker
and dissolve it in de-ionized water. Then, pour this
solution into 1-liter volumetric flask and add de-ionized
water until the indication line.
Cell constant
Electrode
model
KCl
standard
solution
KCl
Weight
Solution
temp.
Conductivity
value
SI units
100 m
-1
(former unit
designation
1 cm
-1
)
9382 -10D 0.01
mol/L
0.7440 g 0ºC
77.4 mS/m
(0.774 mS/cm)
18ºC
122.0 mS/m
(1.220 mS/cm)
25ºC
140.8 mS/m
(1.408 mS/cm)
Содержание D-25
Страница 2: ......
Страница 11: ...CONTENTS VIII HORIBA ...
Страница 31: ...1 Overview of the Meter 1 9 Using the protective cap D 53 54 55 20 HORIBA ...
Страница 75: ...2 Taking Measurements 2 9 Dissolved oxygen DO measurement D 55 64 HORIBA ...
Страница 139: ...4 RS 232C communications 4 5 Communication example using the HyperTerminal 128 HORIBA ...