6-6
OLYMPIC SMART SCALE
Model 50/51 Instruction Manual
Repair Procedures
Internal Fuse
Required parts:
Pico fuse, x1 (part no. 200021)
ESD protection
Allen wrench
Solder gun and solder
To replace the pico fuse:
1
If the scale is on, press
ON/OFF
to turn off the scale, then unplug the
power cord.
2
Disassemble the scale (see page 6-2).
3
Remove the pico fuse:
a
Remove the power supply PCB (see
page 6-8); the pico fuse is located on
the power supply PCB.
b
Carefully unsolder the fuse, then remove the fuse.
4
Insert the new fuse: Solder the new pico fuse into position F1 on the power
supply PCB.
5
Reassemble and calibrate the scale (see page 6-4, then page 4-3).