OLYMPIC SMART SCALE
Model 50/51 Instruction Manual
6-9
Repair Procedures
If so, test the display PCB ribbon cable; replace if damaged.
7
Continue testing internal components, or reassemble and calibrate the scale (see
Testing the Ribbon Cables
If the scale is not operating correctly test the ribbon cables, which are connected to
the printed circuit boards, for continuity, open shorts, and mis-wires.
Required Items:
ESD protection
To test the PCB ribbon cables:
1
If the scale is on, press
ON/OFF
to turn off the scale, then unplug the
power cord.
2
Disassemble the scale (see page 6-2).
3
Remove the cable and check each wire for continuity.
4
Replace any damaged PCB ribbon cables.
5
Continue testing internal components, or reassemble and calibrate the scale (see
Testing the Main Printed Circuit Board
If the scale lights do not illuminate or appears to be malfunctioning, a PCB may
be damaged. Test the PCB to confirm that it is damaged prior to replacing.
Required Items:
ESD protection
Voltmeter
To test the main PCB:
1
If the scale is on, press
ON/OFF
to turn off the scale, then unplug the
power cord.
2
Disassemble the scale (see page 6-2).
NOTE
#
Schematics, wiring diagrams, and PCB drawings are located on pages A-1–A-9.
Before testing the PCB, check all fuses and connectors.
Содержание OLYMPIC SMART SCALE 50
Страница 2: ......
Страница 4: ......
Страница 12: ...1 6 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Description...
Страница 18: ...2 6 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Testing the Device...
Страница 24: ...3 6 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Operating the Smart Scale...
Страница 28: ...4 4 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Calibrating the Smart Scale...
Страница 46: ...6 14 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Returning for Service...
Страница 48: ...7 2 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Ordering Information...
Страница 64: ...A 10 OLYMPIC SMART SCALE Model 50 51 Instruction Manual...