5-2
OLYMPIC SMART SCALE
Model 50/51 Instruction Manual
Incorrect or
inaccurate weight
Incorrect unit of
measurement
Press
KG/LB
to change unit of
measurement.
Object rubbing against
weighing platform or
platform assembly
Check the weighing platform for objects
(e.g., power cord, blankets) rubbing
against it; remove the objects.
Remove the platform assembly and
check for objects (e.g., cables,
enclosure) interfering with the weighing
platform.
Out of calibration
Calibrate the scale (see page 4-3).
Wrong value weight
used for calibration
Use the appropriate value weight when
calibrating the scale (see page 4-3).
Improper calibration
technique
Use proper techniques when calibrating
the scale (see page 4-3).
Electromagnetic
interference
Check for electromagnetic interference
(see page 8-4).
Damaged load cell
Contact Natus Medical (see page 5-4).
Damaged main PCB
Test the main PCB (see
replace if necessary (see page 6-12).
Damaged power
supply PCB
Test the power supply PCB (see
page 5-4); replace if necessary (see
page 6-12).
Dashes “walk”
across display but
scale won’t show
0
Out of calibration
Calibrate the scale (see
Damaged load cell
Contact Natus Medical (see page 5-4).
Signal beyond zero
capture range or
unstable
With the platform on the scale, place the
scale in counts mode (see
The counts shown must be between
9000 and 13500. If it is out of range,
contact Natus Medical (see page 5-4).
With the platform on the scale, place the
scale in counts mode (see
page 6-12). If
the shown counts fluctuate more than
±4, contact Natus Medical (see
page 5-4).
CHARGING
doesn’t
light when scale is
plugged into
electrical outlet
Blown fuse(s)
Replace the fuse(s) (see page 6-5).
Battery pack isn’t
connected
Check the battery pack connection.
Damaged battery
Test the battery pack to determine if it is
at 8.4V when fully charged. If not,
replace the battery pack (see page 6-7).
Damaged main PCB
Test the main PCB (see
replace if necessary (see page 6-12).
Damaged power
supply PCB
Test the power supply PCB (see
page 6-9); replace if necessary
(see page 6-12).
Damaged display PCB
Test the display PCB (see
replace if necessary (see page 6-12).
Table 5.1
Troubleshooting chart, continued
Problem
Probable Cause
Solution
Содержание OLYMPIC SMART SCALE 50
Страница 2: ......
Страница 4: ......
Страница 12: ...1 6 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Description...
Страница 18: ...2 6 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Testing the Device...
Страница 24: ...3 6 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Operating the Smart Scale...
Страница 28: ...4 4 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Calibrating the Smart Scale...
Страница 46: ...6 14 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Returning for Service...
Страница 48: ...7 2 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Ordering Information...
Страница 64: ...A 10 OLYMPIC SMART SCALE Model 50 51 Instruction Manual...