6-10

OLYMPIC SMART SCALE 

Model 50/51 Instruction Manual

Repair Procedures

 

3

Check the supply voltages (see Table 6.2).

4

If any voltages are out of tolerance, replace the main PCB.

5

Continue testing internal components, or reassemble and calibrate the scale (see 

page 6-4, then page 4-3).

Testing the Power Supply Printed Circuit Board

If the scale lights do not illuminate or appears to be malfunctioning, a PCB may 

be damaged. Test the PCB to confirm that it is damaged prior to replacing.

Required Items:

„

ESD protection

„

Voltmeter

To test the power supply PCB:

1

If the scale is on, press 

ON/OFF

 to turn off the scale, then unplug the 

power cord.

2

Disassemble the scale (see page 6-2).

3

Check the supply voltages (see Table 6.3).

4

If any voltages are out of tolerance, replace the power supply PCB.

5

Continue testing internal components, or reassemble and calibrate the scale (see 

page 6-4, then page 4-3).

Table 6.2

Main PCB test points 

ID

Test Point

Expected Value

+5Vdigital

TP1

5.00 ±0.1Vdc

+5Vanalog

TP2

5.00 ±0.1Vdc

-5Vanalog

TP3

-5.00 ±0.5Vdc

GND

TP8

Reference

NOTE

#

An S1 jumper installed on the main PCB (see page 3-15) will disable the 

KG/LB

 button, 

forcing the scale to only display kilograms.

NOTE

#

„

Schematics, wiring diagrams, and PCB drawings are located on pages A-1–A-9.

„

Before testing the PCB, check all fuses and connectors.

Table 6.3

Power supply PCB test points 

ID

Test Point

Expected Value

Vin

J4-1

12.00 ±0.25Vdc

CHG/AC_ON

J3-1

5.00 ±0.25Vdc

GND

TP1

Reference

Vbat

TP2

7.2Vdc nominal

Vin

J4-1

12.00 ±0.25Vdc

Содержание OLYMPIC SMART SCALE 50

Страница 1: ...nual OLYMPIC SMART SCALE Model 50 51 CAUTION Read and be familiar with this manual before installing operating or servicing this device To ensure operator technician and patient safety use only as spe...

Страница 2: ......

Страница 3: ...The information in this manual is subject to change without notice No part of this manual may be photocopied reproduced translated or reduced to any electronic medium without the express written perm...

Страница 4: ......

Страница 5: ...First Use 2 4 Initial Battery Charge 2 4 Calibration Check 2 4 Testing the Device 2 5 Section 3 Operation 3 1 Operating the Smart Scale 3 1 Setting the Caster Brakes 3 2 Adjusting the Platform Ends 3...

Страница 6: ...the Main Printed Circuit Board 6 9 Testing the Power Supply Printed Circuit Board 6 10 Replacing the Power Supply Printed Circuit Board 6 11 Testing the Display Printed Circuit Board 6 11 Replacing t...

Страница 7: ...for use by qualified service technicians Conventions The following conventions are used in this manual Table 1 1 Conventions Convention Description NOTE Notes provide additional information to clarif...

Страница 8: ...mbols Symbol Definition Symbol Definition Alternating current ac voltage Nonionizing electromagnetic radiation Atmospheric pressure Protective earth ground Caution read instructions Rechargeable batte...

Страница 9: ...sily accommodate the patient see Figure 1 2 The scale operates on rechargeable batteries or power from a hospital grade electrical outlet The Model 50 is a table top model while the Model 51 roll arou...

Страница 10: ...scale AVERAGING yellow The scale is computing a four second average of the patient s weight Avoid touching the patient and platform while the AVERAGING light is flashing OVERLOAD The weight on the pl...

Страница 11: ...electronic smoothing filter is activated and each reading six sample average is compared to the previous reading If the reading is within 200 g it is averaged with the previous reading If eight consec...

Страница 12: ...1 6 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Description...

Страница 13: ...of measurement between metric and English kilograms and pounds You can permanently lock the units to measure kilograms only WARNING Electrical shock hazard when the enclosure is open Unplug the power...

Страница 14: ...ess ON OFF to turn on the scale The units are now locked to kilograms 7 Press KG LB to verify that the units are locked to kilograms If locked the display should still show kilograms 8 Place the Locke...

Страница 15: ...insert the mounting screws through the mounting holes in the cart see Figure 2 3 and tighten until the scale is securely attached 4 Remove the backing from the caution label which is included with the...

Страница 16: ...batteries for the first time charge the batteries for approximately 2 5 hours to reach full charge The scale can be used while the batteries are charging For more information see page 4 2 Calibration...

Страница 17: ...weights part no 56357 To test the Smart Scale 1 Press and hold KG LB while pressing ON OFF 2 Verify that all lights illuminate 3 Place the calibration weights on the scale then confirm that the corre...

Страница 18: ...2 6 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Testing the Device...

Страница 19: ...he scale Keep the platform ends in either the full down or full up positions Ends are locked in the full up position only when LATCHED appears see Figure 3 3 Do not allow the patient to sit or stand o...

Страница 20: ...ver to set the brake step on the opposite side to release the brake see Figure 3 1 There are two locking casters on the Model 51 cart Figure 3 1 Setting the caster brake NOTE For information on averag...

Страница 21: ...ee Figure 3 3 Figure 3 2 Adjusting the platform ends Figure 3 3 Locking the ends into position NOTE Lower one or both ends to accommodate the patient being weighed Whenever possible at least one platf...

Страница 22: ...weight is computed the LOCKED light illuminates and the weight locks on the display Weight Lock Unlock AVERAGE UNLOCK When the LOCKED light illuminates the weight remains locked on the display Press A...

Страница 23: ...t locked weight for 10 seconds and then turn off Weighing the Next Patient Before weighing the next patient the scale display must show 0 If the display is locked LOCKED light illuminates press AVERAG...

Страница 24: ...3 6 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Operating the Smart Scale...

Страница 25: ...soap and water solution or the liquid disinfectant then wipe down all exterior surfaces of the Smart Scale Use the second cloth to dry all exterior surfaces of the Smart Scale WARNING Electrical shoc...

Страница 26: ...0 minutes of continuous battery powered operation time remains If rapidly blinking the battery charge has been depleted and the scale will automatically turn off Connect the scale to a functioning ele...

Страница 27: ...hen press ON OFF to enter calibration mode see Figure 4 1 confirm the following The current software version e g CAL 1 0 appears C 0 appears while the scale captures the zero point C 12500 appears pro...

Страница 28: ...4 4 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Calibrating the Smart Scale...

Страница 29: ...wer supply printed circuit board PCB Test the power supply PCB see page 6 9 replace if necessary see page 6 12 Damaged display PCB Test the display PCB see page 6 9 replace if necessary see page 6 12...

Страница 30: ...page 6 12 Dashes walk across display but scale won t show 0 Out of calibration Calibrate the scale see page 4 3 Damaged load cell Contact Natus Medical see page 5 4 Signal beyond zero capture range or...

Страница 31: ...st the main PCB see page 6 9 replace if necessary see page 6 12 Scale won t lock weight on display and LOCKED doesn t illuminate Damaged main PCB Test the main PCB see page 6 9 replace if necessary se...

Страница 32: ...73 Email Seattle_technical_service natus com Error code E905 Check Sum Error Non volatile memory corrupted or damaged Calibrate the scale see page 4 3 Test the main PCB see page 6 9 replace if necessa...

Страница 33: ...Circuit Boards Observe proper handling cautions at all times The integrated circuits and semiconductors on the printed circuit boards PCBs can be damaged by electrostatic discharge or contamination b...

Страница 34: ...cessing the socket head cap screws 4 Lift the platform assembly up to expose the weighing platform see Figure 6 3 WARNING Electrical shock hazard when the enclosure is open Unplug the power cord from...

Страница 35: ...platform a Remove the two socket head cap screws and the two flat head cap screws see Figure 6 2 b Lift the weighing platform to expose the PCBs see Figure 6 3 c Place the spacer in a safe location Fi...

Страница 36: ...ly 90 in lb torque to the two socket head cap screws Tighten the two flat head cap screws firmly these screws require minimal force to secure c Slide one end of the scale so that it hangs over the edg...

Страница 37: ...from the back of the device Main Fuses Required parts Main fuses x2 part no 200021 To replace the main fuses 1 If the scale is on press ON OFF to turn off the scale then unplug the power cord 2 Remov...

Страница 38: ...s on press ON OFF to turn off the scale then unplug the power cord 2 Disassemble the scale see page 6 2 3 Remove the pico fuse a Remove the power supply PCB see page 6 8 the pico fuse is located on th...

Страница 39: ...n electrical outlet The CHARGING light will illuminate 6 Properly dispose of the nickel cadmium Ni Cad battery The internal battery pack can be recycled by contacting the Rechargeable Battery Recyclin...

Страница 40: ...ble connector 4 Verify that the membrane switch panel is properly connected 5 Using two small jumper wires check the resistance of the malfunctioning switch at the corresponding pin on J2 of the displ...

Страница 41: ...ower cord 2 Disassemble the scale see page 6 2 3 Remove the cable and check each wire for continuity 4 Replace any damaged PCB ribbon cables 5 Continue testing internal components or reassemble and ca...

Страница 42: ...ble the scale see page 6 2 3 Check the supply voltages see Table 6 3 4 If any voltages are out of tolerance replace the power supply PCB 5 Continue testing internal components or reassemble and calibr...

Страница 43: ...r supply PCB into position place and tighten the mounting screws then connect the connectors to the power supply PCB 5 Reassemble and calibrate the scale see page 6 4 then page 4 3 Testing the Display...

Страница 44: ...see page 6 4 then page 4 3 Entering Counts Mode If an electrical problem is suspect such as the signal is beyond zero or the system appears unstable enter counts mode To enter counts mode 1 When the s...

Страница 45: ...1 Contact Natus Customer Service for a Return Merchandise Authorization RMA number and the location where the equipment should be sent 2 Clean the device securely package it and include the RMA numbe...

Страница 46: ...6 14 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Returning for Service...

Страница 47: ...tem number to the customer service representative NOTE Only use Natus Medical approved parts with the Smart Scale Table 7 1 Parts and accessories Item Name Description Item Number Battery pack 401391...

Страница 48: ...7 2 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Ordering Information...

Страница 49: ...attery Pack Single pack with six 1700mAh SC Ni Cad rapid charge batteries 7 2Vdc nominal Environmental Temperature Operating 50 104 F 10 40 C Shipping 40 158 F 40 70 C Storage 4 113 F 20 45 C Relative...

Страница 50: ...ing Information on page 7 1 for more information Warranty One year warranty Electromagnetic Compatibility The Smart Scale complies with the requirements of IEC 60601 1 2 2001 For information see pages...

Страница 51: ...floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast transient burst IEC 61000 4 4 2 kV for power supply lines 1 kV for input output lines 1 kV for po...

Страница 52: ...compliance level in each frequency range 2 Interference may occur in the vicinity of equipment marked with the following symbol 1 Field strengths from fixed transmitters such as base stations for radi...

Страница 53: ...mmunications equipment For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequen...

Страница 54: ...8 6 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Electromagnetic Specifications Manufacturer s Declaration...

Страница 55: ...OLYMPIC SMART SCALE Model 50 51 Instruction Manual A 1 A p p e n d i x A Drawings and Schematics Figure A 1 Smart Scale Model 50 51 Power supply PCB assembly Drawing No 400864 Rev C...

Страница 56: ...A 2 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Figure A 2 Smart Scale Model 50 51 Display PCB assembly Drawing No 400866 Rev A...

Страница 57: ...OLYMPIC SMART SCALE Model 50 51 Instruction Manual A 3 Figure A 3 Smart Scale Model 50 51 Main PCB assembly Drawing No 401094 Rev E...

Страница 58: ...A 4 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Figure A 4 Smart Scale Model 50 51 Power supply schematic Drawing No 700141 Rev...

Страница 59: ...OLYMPIC SMART SCALE Model 50 51 Instruction Manual A 5 Figure A 5 Smart Scale Model 50 51 Display schematic Drawing No 700144 Rev A...

Страница 60: ...A 6 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Figure A 6 Smart Scale Model 50 51 Main schematic Drawing No 700152 page 1 Rev F...

Страница 61: ...OLYMPIC SMART SCALE Model 50 51 Instruction Manual A 7 Figure A 7 Smart Scale Model 50 51 Main schematic Drawing No 700152 page 2 Rev F...

Страница 62: ...A 8 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Figure A 8 Smart Scale Model 50 51 Main schematic Drawing No 700152 page 3 Rev F...

Страница 63: ...OLYMPIC SMART SCALE Model 50 51 Instruction Manual A 9 Figure A 9 Smart Scale Model 50 51 Wiring diagram Drawing No 700251 Rev A...

Страница 64: ...A 10 OLYMPIC SMART SCALE Model 50 51 Instruction Manual...

Отзывы: