OLYMPIC SMART SCALE
Model 50/51 Instruction Manual
5-3
CHARGING
lights,
but scale won’t
operate on battery
power
Depleted battery
Plug the device into an electrical outlet
and allow the battery to charge for
2.5 hours. If the battery doesn’t charge,
replace the battery pack (see
Damaged battery
Test the battery pack to determine if it is
at 8.4V when fully charged. If not,
replace the battery pack (see page 6-7).
Power supply PCB
damaged
Replace the power supply PCB (see
LOCK/AVERAGE
won’t lock or average
weight
Damaged membrane
switch panel
Test the membrane switch panel (see
page 6-8).
Damaged display
panel cable
Test the display panel cable (see
Damaged main PCB
Test the main PCB (see
replace if necessary (see page 6-12).
ZERO
won’t zero
object on scale
Objects not on scale
Place the objects (e.g., blankets, toys) on
the scale, then press
ZERO
to cancel
their weight. Next, add the patient to the
scale and read the weight.
Damaged membrane
switch panel
Test the membrane switch panel (see
Damaged display
panel cable
Test the display panel cable (see
Damaged main PCB
Test the main PCB (see
replace if necessary (see page 6-12).
Scale won’t lock
weight on display
and
LOCKED
doesn’t illuminate
Damaged main PCB
Test the main PCB (see
replace if necessary (see page 6-12).
RECALL
won’t
display last locked
weight
Scale display on
RECALL
only operates when the scale
display is off.
Damaged membrane
switch panel
Test the membrane switch panel (see
Damaged display
panel cable
Test the display panel cable (see
Damaged main PCB
Test the main PCB (see
replace if necessary (see page 6-12).
RECALL
always
shows zero
Damaged main PCB
Test the main PCB (see
replace if necessary (see page 6-12).
Scale won’t calibrate;
stops at C 0
Object on platform
Ensure that no objects (e.g., weight) are
on the scale.
Excess vibration
Place the scale on a stable surface.
Damaged load cell
Contact Natus Medical (see page 5-4).
Scale won’t calibrate;
stops at C 12500
Incorrect calibration
weight
Use an NIST traceable 12.5-kg weight.
Excess vibration
Place the scale on a stable surface.
Damaged load cell
Contact Natus Medical (see page 5-4).
Error code E901
(Cal Error)
Counts out of
allowable capture
window
Calibrate the scale (see
Possible load cell failure; contact Natus
Medical (see page 5-4).
Table 5.1
Troubleshooting chart, continued
Problem
Probable Cause
Solution
Содержание OLYMPIC SMART SCALE 50
Страница 2: ......
Страница 4: ......
Страница 12: ...1 6 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Description...
Страница 18: ...2 6 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Testing the Device...
Страница 24: ...3 6 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Operating the Smart Scale...
Страница 28: ...4 4 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Calibrating the Smart Scale...
Страница 46: ...6 14 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Returning for Service...
Страница 48: ...7 2 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Ordering Information...
Страница 64: ...A 10 OLYMPIC SMART SCALE Model 50 51 Instruction Manual...