5-4
OLYMPIC SMART SCALE
Model 50/51 Instruction Manual
Technical Support
Technical Support
For technical support, contact Natus Medical at:
Toll-free: 1-866-940-7143 (US/Canada)
Phone:
+1-206-767-3500 (international)
Fax:
+1-206-767-0573
Email:
Error code E905
(Check Sum Error)
Non-volatile memory
corrupted or damaged
Calibrate the scale (see
Test the main PCB (see
replace if necessary (see page 6-12).
Unreadable
display/garbled
Electrical fast transient
bursts
Install a filter on power input lines.
Contact Natus Medical (see page 5-4).
Table 5.1
Troubleshooting chart, continued
Problem
Probable Cause
Solution
Содержание OLYMPIC SMART SCALE 50
Страница 2: ......
Страница 4: ......
Страница 12: ...1 6 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Description...
Страница 18: ...2 6 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Testing the Device...
Страница 24: ...3 6 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Operating the Smart Scale...
Страница 28: ...4 4 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Calibrating the Smart Scale...
Страница 46: ...6 14 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Returning for Service...
Страница 48: ...7 2 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Ordering Information...
Страница 64: ...A 10 OLYMPIC SMART SCALE Model 50 51 Instruction Manual...