OLYMPIC SMART SCALE
Model 50/51 Instruction Manual
8-3
Electromagnetic Specifications – Manufacturer’s Declaration
Electromagnetic Specifications –
Manufacturer’s Declaration
The Olympic Smart Scale is intended for use in the electromagnetic environment specified below. The
customer or the user of the Smart Scale should assure that it is used in such an environment.
Table 8.1
Electromagnetic emissions
Emissions Test
Compliance
Electromagnetic Environment — Guidance
Radio Frequency (RF)
Emissions
CISPR 11
Group 1
The Smart Scale uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class B
The Smart Scale is suitable for use in all establishments including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
Table 8.2
Electromagnetic immunity
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic Environment — Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete, or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for
input/output lines
±1 kV for power
supply lines
Not applicable
Mains power quality should be that of a typical
commercial or hospital environment. Filters on the
power input lines may be necessary (see
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
±1 kV differential
mode
±2 kV common
mode
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions, and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5%
U
T
(>95% dip
in
U
T
) for 0.5 cycle
40%
U
T
(60% dip in
U
T
) for 5 cycles
70%
U
T
(30% dip in
U
T
) for 25 cycles
<5%
U
T
(>95% dip
in
U
T
) for 5 seconds
<5%
U
T
(>95% dip
in
U
T
) for 0.5 cycle
40%
U
T
(60% dip in
U
T
) for 5 cycles
70%
U
T
(30% dip in
U
T
) for 25 cycles
<5%
U
T
(>95% dip
in
U
T
) for 5 seconds
Mains power quality should be that of a typical
commercial or hospital environment. If the user of
the Smart Scale requires continued operation
during power mains interruptions, it is
recommended that the Smart Scale be powered
from an uninterruptible power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE
#
U
T
is the alternating current (ac) mains voltage prior to the application of the test level.
Содержание OLYMPIC SMART SCALE 50
Страница 2: ......
Страница 4: ......
Страница 12: ...1 6 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Description...
Страница 18: ...2 6 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Testing the Device...
Страница 24: ...3 6 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Operating the Smart Scale...
Страница 28: ...4 4 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Calibrating the Smart Scale...
Страница 46: ...6 14 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Returning for Service...
Страница 48: ...7 2 OLYMPIC SMART SCALE Model 50 51 Instruction Manual Ordering Information...
Страница 64: ...A 10 OLYMPIC SMART SCALE Model 50 51 Instruction Manual...