93
Calibration
GB
SUPREMATouch
Checking the Display
If, in spite of correctly calibrated active transmitters, the expected values (0% of the measuring
range for a signal current of 4 mA and 100% of the measuring range for a signal current of 20 mA)
do not appear on the SUPREMATouch, the calibration on the SUPREMATouch must be checked
and corrected if necessary.
For this purpose, either the signal current of the connected sensor or a variable power source can
be used. If the signal current of the sensor is used, make sure that the sensor is supplying the
correct values.
To correct a possibly incorrectly set measuring point, change the selected type of sensor:
(1) Go to the
Setup/Input and Outputs/Measuring Points/Sensor Data
menu.
(2) Select any other type of sensor.
(3) Confirm the selection with
OK
.
(4) Re-select the type of sensor connected and confirm with
OK
.
The measuring point will be set back again to the standard setting of 4 mA = 0% of the
measuring range and 20 mA = 100% of the measuring range.
NOTICE
During this calibration the measuring point must be inhibited manually (alarm rejection).
NOTICE
Adjustments on the MAI module are neither necessary nor possible for active transmitters.
Calibration with a Variable Power Source
(1) Lock the measuring point in question in the
Setup/Input and Outputs/Measuring points
menu
to prevent triggering an alarm.
(2) Detach cable connections of sensor to MAT module.
(3) Connect the variable power source to the MAT module as follows:
MAT module terminal 1: + pole of the power source (signal)
MAT module terminal 4:–pole of the power source (GND)
(4) Set output current of power source to 4 mA.
(5) Unlock measuring point in question in the
Setup/Input and Outputs/Measuring points
menu
to allow a calibration.
(6) Select
Maintain/Calibration/Standard
menu.
(7) Select measuring point to be calibrated in
Measuring Point
field.
(8) Enter 0 % of measuring range in the field
Zero Gas
as zero gas concentration.
(9) Enter 100 % of measuring range in the field
Test Gas
as test gas concentration.
(10) After tapping the
Start
button, enter the required password or use the key switch.
End Calibration submenu displays.
Current measurement value UA of the measuring point to be calibrated will appear in Sig:
field.
For an input current of 4 mA, a value of 400 mV ± 10 mV should be displayed.
In the field Ux= no value or ***** is displayed.
(11) If value U
A
is within tolerance range (400 mV ± 10 mV), confirm with
Store
button.
The value will be appear in CAL-ZERO.
Содержание SUPREMATouch MCP 20
Страница 1: ...MSAsafety com Operating Manual SUPREMATouch Fire and Gas Warning Unit Order No 10121863 05...
Страница 8: ...MSAsafety com User Instruction Manual SUPREMATouch Fire and Gas Warning Unit...
Страница 72: ...MSAsafety com Service and Maintenance Guide SUPREMATouch Fire and Gas Warning Unit...
Страница 101: ...8645 Rapperswil Jona Switzerland info ch MSAsafety com www MSAsafety com MSA 2017 All rights reserved...
Страница 232: ...232 Dimensions GB SUPREMATouch 18 Dimensions 18 1 Rack...
Страница 233: ...233 Dimensions GB SUPREMATouch 18 2 Rail mounted Modules MRO8 TS Module MRO16 TS Module...
Страница 235: ...235 Dimensions GB SUPREMATouch MRO20 8 TS Module 1 3 2 69 90...
Страница 236: ...236 Dimensions GB SUPREMATouch MRO20 16 TS Module 2 5 2 64 73 relay dependent 90...
Страница 237: ...237 Dimensions GB SUPREMATouch MRC TS Module MGT40 TS Module...
Страница 238: ...238 Dimensions GB SUPREMATouch MHD TS Module MAT TS Module...
Страница 241: ...For local MSA contacts please visit us at MSAsafety com Because every life has a purpose...