91
Calibration
GB
SUPREMATouch
1-Man Calibration
For a one-man calibration, the person must perform the following steps:
(1) Select
Maintain/Calibration/Standard
menu.
(2) Select input to be calibrated in “Measuring Point” field
(3) Enter gas concentration in
Zero Gas
field:
a) Enter the concentration of the test gas in the zero gas (usually 0%), not the concentration
of the zero gas.
(4) Enter test gas concentration in
Test Gas
field.
(5) If test gas used is different from reference gas entered in the
Setup/Inputs&Outputs/
Measuring points
menu, change the entry in field
Test Gas
of the
Calibration
submenu.
(6) Start calibration with
Start 1-Man
button.
a) Confirm messages immediately shown on the screen.
(7) Go to the sensor in the field.
(8) Supply zero gas via test adapter to the sensor assigned to the selected measuring point for
a sufficient period of time (duration and flow rate according to sensor operating and mainte-
nance instructions).
(9) Stop zero gas supply and supply test for a sufficient period of time.
(10) Stop test gas supply and return to the SUPREMA.
The measurement values taken over during the calibration process is shown in the NEW line.
(11) Check the measuring values taken over and shown on the display.
If plausible, complete the calibration by pressing the “End” button; otherwise cancel the cali-
bration by pressing the “Cancel” button.
WARNING!
If the signal voltage exceeds 2000 mV or is too weak during the test gas supply, the calibration is
invalid. Under no circumstances may the calibration value be accepted. Check test gas concen-
tration and make sure it is being supplied correctly. It may be necessary to check and correct the
preadjustment of the measuring point.
If the preadjustment was correct, the ACTUAL VALUES for the zero point will be approx-
imately in the range of 350 mV–450 mV. The signal voltage shown is calculated
according to the formula: Signal = C / 100 * 1600 mV + 400 mV (for sensors with a linear
output signal), where C is the concentration of the test gas in % of the measuring range.
The tolerance is approximately equal to the signal value in mV ± 100 mV.
If the preadjustment was correct, the ACTUAL VALUES for the zero point will be approx-
imately in the range of 350 mV–450 mV. The signal voltage shown is calculated
according to the formula: Signal = C / 100 * 1600 mV + 400 mV (for sensors with a linear
output signal), where C is the concentration of the test gas in % of the measuring range.
The tolerance is approximately equal to the signal value in mV ± 100 mV.
Содержание SUPREMATouch MCP 20
Страница 1: ...MSAsafety com Operating Manual SUPREMATouch Fire and Gas Warning Unit Order No 10121863 05...
Страница 8: ...MSAsafety com User Instruction Manual SUPREMATouch Fire and Gas Warning Unit...
Страница 72: ...MSAsafety com Service and Maintenance Guide SUPREMATouch Fire and Gas Warning Unit...
Страница 101: ...8645 Rapperswil Jona Switzerland info ch MSAsafety com www MSAsafety com MSA 2017 All rights reserved...
Страница 232: ...232 Dimensions GB SUPREMATouch 18 Dimensions 18 1 Rack...
Страница 233: ...233 Dimensions GB SUPREMATouch 18 2 Rail mounted Modules MRO8 TS Module MRO16 TS Module...
Страница 235: ...235 Dimensions GB SUPREMATouch MRO20 8 TS Module 1 3 2 69 90...
Страница 236: ...236 Dimensions GB SUPREMATouch MRO20 16 TS Module 2 5 2 64 73 relay dependent 90...
Страница 237: ...237 Dimensions GB SUPREMATouch MRC TS Module MGT40 TS Module...
Страница 238: ...238 Dimensions GB SUPREMATouch MHD TS Module MAT TS Module...
Страница 241: ...For local MSA contacts please visit us at MSAsafety com Because every life has a purpose...