Adjustment of the average-time for CO2 NDIR-bench (optional)
If the analyser is equipped with a CO2/CH4 NDIR-bench, the user has the possibility to set the
average time.
1. Open the path: Extras/ Adjustment menu/ NDIR CO2/CH4 adjustment.
2. Open “averaging” with the
F1
-key.
3. Configuration of averaging time: In this menu is able to set the averaging time with the
left/
right arrow keys
and the
F1
or
F3
keys.
The fast averaging time
: 20 s
The smooth averaging time
: 240 s
NOTE
Effect of the average time
A smooth averaging time smooths the measurement signal but increase
the response time of the measurement signal.