![mru SWG100 CEM Скачать руководство пользователя страница 130](http://html.mh-extra.com/html/mru/swg100-cem/swg100-cem_user-manual_1846047130.webp)
MRU GmbH
SWG 100 User Manual
Page
130
of
155
Bending radius of heated sampling lines
The picture below shows the minimal bending radius.
ATTENTION
Falling below the bending radius will reduce life time of the sampling
probe.
Mounting rules for heated sampling lines
Basic rules for laying the heated sampling lines
Wrong
Right
Bending radius min. = 32 cm
Do not lay several heated sampling lines
close together. -> Overheating hazard!!
Care for enough space between the
single heated sampling lines.
Do not use clips, being too tight. ->
Crushing hazard!!
Only use clips, which have the right
side.