5.10.
Connection of main power supply
WARNING
Electrical shock!
Electricity could cause damage, injuries and even death! Only educated
staff should install the device.
The analyser needs a main power supply of
100-230 VAC/ 47-60 Hz
.
The power supply will be connected on the hat-rail.
Exactly:
•
L and N phase on the circuit breaker (
A
in sketch below).
•
PE on the connection terminal (
B
in sketch below.)
100-230 Vac/ 47-60 Hz/ 300W
N
L
PE
Position of the hat-rail. On the hat-rail can be found:
•
Main power supply.
•
I/O moduls (max. 10, 1 basic, all other optional)
•
H2 TCD (optional).
A
B
N
L
PE
Pre-fuse