MRU GmbH
SWG 100 User Manual
Page
83
of
155
Measurement SPX (Cycle phase details):
In the cycle phase details of “Sample point X” the measurement time
and the suction delay can be configured. Each sample point can be configured individually. In the cycle phase
details the following times can be set:
MEASUREMENT SPx
Measuring site valves
Valve of selected site is open, others closed
Zeroing valve
Valve closed
Duration
Phase duration: 2 min. to 24 h
Suction/response time: 30 sec. to 1h
Pure measurement: calculated
H2S-low: Activated/protect (Optional)
Stand-by (cycle phase details):
In the cycle phase details “Stand
-
by” the sleep mode time can be configured. In
the cycle phase details the following times can be set:
•
Phase duration:
Entire Stand-by time (Purging time + Quiet time = Phase time).
•
Purging time:
The time, to purge the analyser with ambient air, through the zero gas inlet.
•
Quiet time:
The time, where the analyser is in the pure stand-by mode.
STAND-BY
Measuring site valves
Valves closed
Zeroing valve
Valve closed
Duration
Phase duration: 2 min to 24h
Purging time: 30 sec. to 1h
Quiet time: calculated
•
Purging (cycle phase details):
The purging is a separate configuration point to purge the analyser with ambient
air through the zero gas inlet. It can be helpful, if the analyser must switch between a sample point with different
sample gas concentrations.
Содержание SWG100 CEM
Страница 1: ...SWG100 User Manual MRU GmbH D 74172 Neckarsulm ...
Страница 2: ...MRU GmbH SWG 100 User Manual MRU GmbH D 74172 Neckarsulm 2 155 ...
Страница 98: ...MRU GmbH SWG 100 User Manual Page 98 of 155 Hardware state tests 1 2 3 4 5 6 7 8 9 2 3 4 5 6 7 8 9 1 ...
Страница 102: ......
Страница 135: ...MRU GmbH SWG 100 User Manual Page 135 of 155 13 3 Option O2 paramagnetic cell measurement 0 25 00 ...
Страница 142: ...MRU GmbH SWG 100 User Manual Page 142 of 155 13 10 Option NDIR bench for CO2 analysis 0 100 ...
Страница 154: ...MRU GmbH SWG 100 User Manual Page 154 of 155 14 Declaration of conformity ...
Страница 155: ...MRU GmbH SWG 100 User Manual Page 155 of 155 V1 00 EN ...