5.3.
Installation of the analyzer
The analyzer is equipped with adapters for the wall mounting. The dimensions for the wall mounting are shown in
picture below.
Picture 1: Dimensions for the wall mounting.
CAUTION
Only operate the analyzer in an upright position!
Only power the device up after it is correctly mounted!
General installation rules
•
Mount the device on a solid wall or steel rack.
•
Be sure, that the air circulation is not obstructed.
•
Let enough room for the tubing or piping.
For outdoor installation
Ensure that the analyzer is mounted on a rain and sun protected place (weather shade).
For indoor installation:
Ensure that the analyzer is installed on a dry and clean place. Be sure that the room is permanently vented with
fresh air.
Connect the VENT gas-outlet of the device to ambient by using adequate tube with min Ø8mm ID.