9.4.
Update the firmware
The analyzer and different installed options can be updated, if it is necessary. Following options
can be updated:
•
The firmware from the analyzer.
•
The firmware from the pcb- mainboard.
•
The firmware from the NDIR-bench.
•
The firmware from the installed I/O modules.
General steps for the firmware-update
1.
Copy the actual firmware for the analyzer or the firmware from the option on a SD-
card. Be sure, that the firmware is in the mean root of the SD-card. All firmware
updates have the ending “fwb”.
2.
Put the SD-card on the card slot on the operation unit. The card slot can be found
inside the door (see sketch below).
3.
If the SD-card is recognized, the analyzer will make a noise.
4.
Open the path:
E
XTRA
/D
EVICE
I
NFO
. Dependent from the firmware update it can be
necessary to open the different submenus.
Update the analyzer (Firmware-
Updates with filename “1106.fwb”)
1.
Open the path:
E
XTRA
/D
EVICE
I
NFO
2.
Press
F2 = details
to open the details for the main device menu.
BEEP!!
Содержание SWG100 CEM
Страница 1: ...SWG100 User Manual MRU GmbH D 74172 Neckarsulm ...
Страница 2: ...MRU GmbH SWG 100 User Manual MRU GmbH D 74172 Neckarsulm 2 155 ...
Страница 98: ...MRU GmbH SWG 100 User Manual Page 98 of 155 Hardware state tests 1 2 3 4 5 6 7 8 9 2 3 4 5 6 7 8 9 1 ...
Страница 102: ......
Страница 135: ...MRU GmbH SWG 100 User Manual Page 135 of 155 13 3 Option O2 paramagnetic cell measurement 0 25 00 ...
Страница 142: ...MRU GmbH SWG 100 User Manual Page 142 of 155 13 10 Option NDIR bench for CO2 analysis 0 100 ...
Страница 154: ...MRU GmbH SWG 100 User Manual Page 154 of 155 14 Declaration of conformity ...
Страница 155: ...MRU GmbH SWG 100 User Manual Page 155 of 155 V1 00 EN ...